Ways to Reduce Noise Checklist |
||||
| LOCK-IN
AMPLIFIER METHODS
_____ Increase the averaging time carefully, without overloading the lock-in input. See Phase Shifts and Overloading _____ Increase the current through the sample in order to maximize the signal to noise ratio, but be sure not to heat the sample. See Current _____ Try a different oscillator frequency in an attempt to find lower interference. See Current _____ For quasi-1D charge density wave, spin density wave, use a higher frequency because such systems respond differently to low frequency signals. See Current _____ Very high resistance samples need a lower frequency, because cable and other reactance begins to have a higher effect at higher frequencies. See Current _____ One of the biggest problems is overload of the lock-in. If this happens, try decreasing the lock-in sensitivity. See Phase Shifts and Overloading
MECHANICAL DAMPING TO REDUCE VIBRATION-INDUCED SIGNALS IN LARGE FIELDS: _____Bring helium out of superfluid state. Normal-state helium damps out vibrations better than superfluid helium. See Mechanical Noise _____Fasten probe wires over probe with GE varnish for as much of their length as possible, especially near the sample. See Mechanical Noise CAPACITIVE/INDUCTIVE INTERFERENCE REDUCTION METHODS _____ Pay careful attention to grounds and voltage drops along small-gauge shields on ground returns. See Ground Loops _____Unplug all unnecessary electronics before the pulse. including the LR-700, and any computers that may have been used for cool down curves. Also move wires away from the magnet, and other potential noise sources. See External Noise and Coupling Noise _____Filter with Stanford SR560 pre-amp or design own filters. See Filtering
NOISE REDUCTION METHODS _____1/f noise is decreased at higher f. This only becomes important if you have minimized all interference sources and have already obtained a very high signal to noise ratio. Typical 1/f knees can be as low as a few hertz or as high as 40kHz in some ultra-low-noise op-amps. See Intrinsic Noise
SAMPLE MOUNTING _____Excessive contact resistance can degrade signal to noise ratio, this is especially true for very low resistance samples. Be sure sample is mechanically constrained and that thermal contact is correct for the measurement. See Sample Contact Resistance
SOFTWARE METHODS _____ Limiting, averaging and band-pass filtering can often be achieved via software. |
||||